Figure 10-6
The Zero Offset, which sets the channel location of zero keV, ranges from a value of 0 to
255. To increase the Zero offset value, select the right arrow key from the arrow bar located
beneath the Zero Offset value. To decrease the Zero Offset, select the left arrow key from
the arrow bar. For large incremental changes, place the trackball indicator on the square
located in the middle of the arrow bar, and depress the left clicker. With the clicker
depressed, slide the box to the desired setting. Moving the box to the right increases the
Zero Offset value, moving the box to the left decreases the Zero Offset value.
The Threshold value acts as a coarse LLD control, and the LLD value acts as a fine LLD
control adjustment. Together, these values set the lower channel limit of detection for the
MCA board. Both may be adjusted from 0 to 255 by using the right arrow to increment, and
the left arrow to decrement. The LLD default is set to eliminate any extraneous electronic
noise yet allow counts of the lowest possible energy to be detected. The ULD control sets
the upper channel limit of detection. Similarly, the ULD ranges from 0 to 255, and is adjusted
by choosing the right arrow to increase the value and the left arrow to decrease the value.
For large incremental changes, place the trackball indicator on the square located in the
middle of the arrow bar, and depress the left clicker. With the clicker depressed, slide the
box to the desired setting. Moving the box to the right increases the value, moving the box to
the left decreases the value. When the Zero offset, LLD, and ULD values are correct, select
the
CLOSE
button. This will cause the changes to be temporarily stored in the software
memory while in the MCA mode only, and exit the screen.
NOTE: WHEN YOU EXIT THE MCA PROGRAM, AND RETURN TO AN APPLICATIONS
PROGRAM (THYROID UPTAKE, BIOASSAY, WIPE TEST, LAB TESTS), THE
SOFTWARE WILL RESTORE THE HIGH VOLTAGE AND GAIN SETTINGS THAT WERE
DETERMINED IN AUTOCALIBRATE. THIS WILL RETURN THE CALIBRATION TO 2
keV/CHANNEL, AND ALLOW THE AUTOMATIC PEAK FINDING AND ROI PROGRAMS
TO OPERATE CORRECTLY.
Quick Calibration Restore
If changes have been made to any of the calibration parameters, voltage, gain, zero offset,
the automatic energy calibration settings can be quickly restored by selecting
Restore
Calibrated Settings
from the
Setup
list. Figure 10-7 will appear.
July 00
MCA
10-5
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......