CAPINTEC, INC
CAPTUS
®
2000
Figure 5-1
Energy Calibration
The first step in the Autocalibrate Sequence is measurement of the Cs137 standard. This
function uses the 662 keV peak and the 32 keV peak of Cs137 to automatically adjust the
high voltage, coarse gain, fine gain, and zero offset to provide a calibration of approximately
2.0 keV per channel in the MCA.
To start the autocalibration process, click the
AUTOCALIBRATE
button with the track ball or
press (Alt-A). A pop up screen will prompt the user to place the Cs137 standard in position
as shown in Figure 5-2.
5-2
QUALITY ASSURANCE
July 00
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......