Before beginning acquisition, first define the required counting parameters. Choose
Setup
Tests
from the upper menu bar, and then select the
Setup Tests
option. Figure 10-21 will
appear.
Figure 10-21
Enter the length of time for each count in seconds, from 1 to 31999. The software will
automatically count the time as live time to compensate for any changes in count rate
throughout the counting sequence. Next enter the number of repetitions required. The
repetition range is 1 to 1000. Third, enter the counting interval in real time seconds, from 1
to 32000.
Note that for high count rates, the real time interval must be sufficiently long to allow for the
dead time compensation during acquisition. To check the live time and real time required,
count the highest activity expected during the time activity analysis in the MCA mode. The
live time and real time will appear in the upper right corner of the MCA screen. Use these
values to calculate a suitable real time counting interval. Divide the real time by the live time,
and use the resulting value to approximate the minimum required real time counting interval.
For example, the live time counted was 60 seconds and the real time which elapsed was 120
seconds.
2
=
60
120
=
Correction
This indicates that the counting interval in real time must be double (x2) the preset live time.
When all of the parameters have been correctly entered, select the
DONE
button. This will
store the counting parameters in the software memory and return to the Time Activity screen.
The counting sequence information will appear in the results display area, and on any printed
10-18 MCA July
00
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......