![Capintec CAPTUS 2000 Owner'S Manual Download Page 170](http://html1.mh-extra.com/html/capintec/captus-2000/captus-2000_owners-manual_494180170.webp)
CAPINTEC, INC
CAPTUS
®
2000
Figure 9-13
Note that all bioassay measurements must be performed in a low background area to insure
adequate detection of low activity.
Select
OK
to proceed with the background measurement. During the counting process,
COUNTING BACKGROUND will flash in red above the spectral display. Once the
measurement has been completed, Figure 9-14 will appear and require data validation.
Figure 9-14
To accept the data and proceed to the next step in the measurement sequence, select
YES
.
To reject the data and return to the previous counting prompt, select
NO
.
Once the background measurement has been completed and the data saved, Figure 9-15
will appear for the next step in counting sequence, the measurement of the staff member’s
thyroid. Place the collimator flush against the person’s neck to insure maximum sensitivity
and shortest source to detector distance. This provides a source to detector distance of 16.5
cm for the standard ANSI Thyroid Uptake Collimator, and a distance of 6 cm when using the
specially designed Bioassay Collimator. Center the thyroid area in the collimated field of
view.
9-12 BIOASSAY
July
00
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......