CAPINTEC, INC
CAPTUS
®
2000
Figure 4-13
Enter the calibrated activity, calibration date, and calibration time for the source. The
software will automatically decay correct the source to the current activity. To exit the screen
without saving the source information, select
CANCEL
. To save the source information,
select the
DONE
button.
Once the source information has been determined, begin the efficiency measurement
process by selecting the
COUNT BACKGROUND
button. Insure that no sources are near
the detector. The pop up screen shown in Figure 4-14 will appear and request counting time.
Figure 4-14
Enter the appropriate counting time. The system default is 120 seconds. To exit without
measuring select
CANCEL
. Confirm the time by selecting
OK
. Then the system will
automatically begin the background count. To stop the count and save the counting data
before the preset live time has elapsed, choose the
STOP & SAVE
button.
July 00
SYSTEM SETUP
4-13
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......