To proceed with the Co58 standard count, place the Co58 standard sample in the well
counter and select the
OK
button. To cancel the procedure, select the
Cancel
button, and
Figure 8-25 will appear. Once the count is completed, Figure 8-24 will appear and require
confirmation of the data.
Note that during the count acquisition of the Co58 standard, two Regions of Interest (ROI’s)
are highlighted in red. The left ROI is the Co57 region, the area to the right is the Co58 ROI.
The software is counting the spillover counts from the Co58 standard which appear in the
Co57 ROI. This spillover ratio is automatically calculated by the software, and the Co57
activity found in the patient’s urine sample will be properly corrected.
Once the standards have been counted and data accepted and saved, the next step in the
counting sequence is the measurement of the patient’s urine. Figure 8-29 will appear.
Figure 8-29
To proceed with the patient’s urine count, place the urine sample in the well counter and
select the
OK
button. To cancel the procedure, select the
Cancel
button, and Figure 8-25
will appear. After the count is completed, Figure 8-24 will appear and require confirmation of
the data.
Once the urine count has been completed and the data accepted and saved, Figure 8-22
will appear again. PROCEDURE COMPLETE will appear in blue at the bottom of the
spectral display area, and only the
EXIT
button will be active. Choose the
EXIT
button to
return to the DICOPAC Test Main screen.
8-22
LAB TESTS
July 00
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......