COUNTING TIME AND PROBE DISTANCE
From the main THYROID UPTAKE screen as seen in Figure 6-1, select the
Setup
command
from the menu bar at the top left. Select the
Defaults
listing that appears in the drop down
list. A Default Settings window will appear as seen in Figure 6-14.
Figure 6-14
Enter the desired counting time and probe distance as shown in Figure 6-14. Default
settings are 300 seconds and a probe distance of 25 cm. Changing them creates new
defaults. By selecting a probe distance between 24.5 cm and 35 cm, the field of view will
meet ANSI 44.3. The counting time should be set to provide at least 10,000 total counts.
Note that the entire counting time need not elapse for each measurement. During
acquisition, a
STOP & SAVE
button is available to stop the count before the preset time has
elapsed and save the count data. The counting time and probe distance will appear in each
patient demographic screen. These parameters may be changed for any patient as required.
After changing the probe distance and counting time default settings, select
DONE
to save
them or
CANCEL
to return to the main THYROID UPTAKE screen.
ADDING NEW PATIENTS
From the main THYROID UPTAKE screen, Figure 6-1, choose the
ADD
button. The Edit
Patient Library screen will appear as shown in Figure 6-15. Enter your patient demographic
data: patient name, ID, date of birth, and physician. Move from one data entry field to
another by using the
Tab
key, or placing the cursor in the appropriate field and clicking the
trackball. The date of birth, middle initial, and physician are optional fields and may be left
blank. The ID may be any combination of 16 alphanumeric characters.
July 00
THYROID UPTAKE
6-11
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......