setup information and return to the main autocalibration screen, or
CANCEL
to exit without
saving and return to the main autocalibration screen. The Constancy Test uses this
calibration information to decay correct the Cs137 rod source.
Select the
CONSTANCY
button from the main Autocalibrate screen which is shown in Figure
5-1. Figure 5-6, will appear.
Figure 5-6
Select
OK
to begin the Cs137 measurement process, or
Cancel
to return to the main
autocalibration screen. Since the constancy test is a measure of reproducibility, it is very
important that the source be placed in exactly the same position each time the constancy test
is performed. Place the Cs137 rod source in the well of a drilled well detector with the black
tip down.
For an easy way to insure reproducible source placement for the uptake probe, turn the
uptake probe so that the collimator opening is facing upwards towards the ceiling. Gently
place the black, flat plastic end of the source onto the middle of the surface of the crystal.
NOTE:
The source is plastic and can be safely placed against the surface of the detectors
when handled carefully. This placement will provide good positioning for reproducibility and
acceptable count rates for quick calibration of the flat field detector using the reference
source provided with the system.
Once the source is correctly placed, select
OK
to begin the Reproducibility Test, or
CANCEL
to return to the main autocalibration screen. To stop the counting process, select
ABORT
during acquisition. You will return to the main autocalibration screen. When the process is
complete, "DONE" appears in the MCA display area. The system will count until at least
2000 vertical scale peak counts have been acquired. After completion of the constancy test,
the total count rate, live count time, calculated standard activity, measured activity, and
percent error will appear in the display area. The results of each constancy test is
automatically saved to a calibration file. The daily quality assurance tests should only take
about three minutes for each detector.
5-6
QUALITY ASSURANCE
July 00
Summary of Contents for CAPTUS 2000
Page 2: ......
Page 8: ......
Page 26: ......
Page 47: ...CAPINTEC INC CAPTUS 2000 Figure 4 16 July 00 SYSTEM SETUP 4 15...
Page 48: ...CAPINTEC INC CAPTUS 2000 4 16 SYSTEM SETUP July 00...
Page 52: ......
Page 63: ...CAPINTEC INC CAPTUS 2000 Figure 5 9 July 00 QUALITY ASSURANCE 5 11...
Page 66: ......
Page 105: ...CAPINTEC INC CAPTUS 2000 Figure 7 10 July 00 WIPE TESTS 7 11...
Page 133: ...CAPINTEC INC CAPTUS 2000 Figure 8 31 July 00 LAB TESTS 8 25...
Page 145: ...CAPINTEC INC CAPTUS 2000 Figure 8 46 July 00 LAB TESTS 8 37...
Page 157: ...CAPINTEC INC CAPTUS 2000 July 00 LAB TESTS 8 49 Figure 8 59...
Page 158: ......
Page 165: ...CAPINTEC INC CAPTUS 2000 Figure 9 7 July 00 BIOASSAY 9 7...
Page 173: ...CAPINTEC INC CAPTUS 2000 July 00 BIOASSAY 9 15 Figure 9 17...
Page 174: ......
Page 190: ...CAPINTEC INC CAPTUS 2000 Figure 10 19 10 16 MCA July 00...
Page 198: ...CAPINTEC INC CAPTUS 2000 10 24 MCA July 00 Figure 10 30...
Page 204: ...CAPINTEC INC CAPTUS 2000 11 6 CLEANING AND MAINTENANCE July 00 Figure 11 2...
Page 206: ......