40 | Chapter 3: Operation
Model
Buffer Switching One Buffer
405 TS/LS
300
250
HT
300
250
Buffer Switching models have four buffer valves; otherwise, only one reagent bottle
may be connected to the washer. For Select models (with Cell Wash (CW)
capability), the recommended prime volume is the same with and without Buffer
Switching.
Model Quick Prime
Prime step
Select
300
Main
Low Flow
150
150
Manifold Prime
: 50 mL is recommended to clear air from the manifold only and to
wet the tips. This is recommended before running a protocol to correct fluid loss
due to evaporation.
Dead Volume
BioTek's recommended prime volumes are based on purity testing and measured
dead volumes. Generally, priming with three times the dead volume assures
purity.
Model
Dead Volume
405 TS/LS
108 mL
HT
108 mL
Select
129 mL
For models with buffer switching tubing, dead volume is ± 2 mL additionally.
Optimize protocols to improve evacuation
When a wash protocol leaves too much residual fluid in the wells, optimize the
protocol with these recommendations:
l
Add a secondary aspiration to a wash cycle on page 70, including
Final Aspirate
,
l
Decrease the aspirate
Travel Rate
,
l
Add a
Delay
to the Aspirate and/or Final Aspirate step ,
l
Lower the aspirate height (Z-axis position).
405™ Microplate Washer LS
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......