AutoPrime | 123
LHC:
Keypad:
4. Click
Send
to transfer the
settings to the instrument.
5. Define the AutoPrime
parameters: rate, volume, and
buffer valve, if applicable. Press
Enter at each screen to advance
to the next.
AutoPrime for Overnight-Weekend Maintenance
AutoPrime can be used to keep the manifold tubes wetted during idle periods
throughout the day and then modified to soak the tubing for longer periods
overnight and on the weekends.
See AutoPrime on previous page
for setup
instructions.
AutoPrime Parameters
Washer
AutoPrime Interval
15 minutes
Device
Washer
AutoPrime
Yes/Enabled
Volume
90 mL
Rate
09
Buffer
Any
Submerge Duration
120 minutes/2 hours
Turn on AutoPrime for the Washer
l
Learn about
AutoPrime
, if you haven't already done so.
n
Keypad users
: Select
UTIL
at the main menu, then
AUTPRM
and specify
parameters.
1. Select
Tools>Instrument Utilities>AutoPrime
.
2.
Fill the
Enabled
checkbox.
3. Set the
AutoPrime Interval
: enter the downtime interval that will trigger an
AutoPrime.
4. Select the
Buffer
to use if you have Buffer Switching installed.
5. Set the
Volume
and
Rate.
The default values are 40 mL at flow rate 9.
BioTek Instruments, Inc.
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......