Decontaminate Exterior Surfaces | 139
70% isopropyl alcohol (as a bleach alternative)
Deionized or distilled water
Priming plate
Safety glasses
Surgical mask
Protective gloves
Lab coat
Biohazard trash bags
Clean cotton cloths
Step-by-Step Decontamination Instructions:
l
Decontaminate Exterior Surfaces below
l
Decontaminate Tubing and Manifold on next page
l
Alternate Decontamination Procedure for Tubing and Manifold on page 141
Decontaminate Exterior Surfaces
Caution!
Be sure to check the percentage NaClO of the bleach you are using; this
information is printed on the side of the bottle. Commercial bleach is typically 10%
NaClO; in this case, prepare a 1:20 dilution. Household bleach is typically 5% NaClO; in
this case, prepare a 1:10 dilution.
The bleach solution is caustic; wear gloves and eye protection when handling.
1. Turn off the instrument and disconnect the power cord. Empty the waste bottle.
2. Prepare an aqueous solution of 0.5% sodium hypochlorite (NaClO, or bleach).
As an alternative, 70% isopropyl alcohol (or 70% ethanol) may be used if the
effects of bleach are a concern.
n
Isopropyl alcohol
is not recommended for removing
proteins
(such
as bovine serum albumin).
3. Moisten a cloth with the bleach solution or alcohol.
Do not soak the cloth
.
l
Wipe the keypad (do not soak). Wipe again with a clean cloth moistened with
deionized or distilled water. Dry immediately with a clean, dry cloth.
l
Remove the mist shield if it is attached. Wipe the inside and outside surfaces of the
mist shield.
BioTek Instruments, Inc.
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......