Verify Performance | 31
Run the Verify test:
LHC
1. Select
Tools>Instrument Utilities>Verify Manifold
and click
Run
a new Test.
l
Buffer
: Select the bottle to use, if applicable.
l
Prime
: optionally, prime the manifold tubes to correct for evaporation loss.
40 mL of fluid is dispensed.
l
%CV Threshold
: optionally, change the %CV to match your lab's standard for
QC tests: 5-15%. The default value of 5% is recommended.
2. Press
Start
to run the test.
3. Assess the results:
Verify Results
.
Do not discard the
Verify Test Plate
. Reuse it for as long as possible.
See
Replacement Procedure for Verify Test Plate on page 134
if it is damaged.
Contact BioTek TAC if any of the tests fail.