192 | Chapter A: Error Codes
System Error Codes
Most of these error conditions require technical expertise to correct. Error code 306
and few other exceptions to this rule are listed with remedies in the
Troubleshooting
section. A few other errors may be caused by an obvious
obstruction to a device's movement or insufficient fluid in a supply vessel. Fix
these kinds of errors and restart your instrument to give it an opportunity to clear
the error code.
Contact BioTek Technical Assistance Center (TAC)
for assistance.
Code
Message
What to do
100
Task was aborted
Restart instrument if this message is
unexpected.
210, 220 Carrier X motor didn’t find
home opto sensor transition
Carrier X motor didn’t find
autocal jig opto sensor
transition
Clean the plate carrier, rails, and glide
strips, using mild detergent and hot water,
70% isopropyl alcohol or ethanol. Restart
the instrument. If the error occurs again,
contact BioTek TAC.
211, 221 Carrier Y motor didn’t find home
opto sensor transition
Carrier Y motor didn’t find
autocal jig opto sensor
transition
Run self test. If error reoccurs, contact
BioTek TAC.
212, 222 Dispense head motor didn’t find
home opto sensor transition,
Dispense head motor didn’t find
autocal jig opto sensor
transition
Run self test. If error reoccurs, contact
BioTek TAC.
213, 223 Wash head motor didn’t find
home opto sensor transition,
Wash head motor didn’t find
autocal jig opto sensor
transition
Run self test. If error reoccurs, contact
BioTek TAC.
220
Carrier X motor didn’t find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.
221
Carrier Y motor didn’t find
autocal jig optical sensor
transition.
Service Only.
Contact BioTek TAC.
405™ Microplate Washer LS
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......