Verify Performance | 75
Verify™ Technology Flowchart Diagram
Note:
When the manifold tubes are clog-free, it is expected that all five tests pass and
normal operation can be resumed.
Verify Test Corrective Action
Corrective Action
1
Clean the Manifold
: begin with AutoClean/Quick
Clean and rerun. If test fails again, remove the
manifold to clean it more thoroughly.
See AutoClean the
Washer on page 125
2
If the test fails after the manifold has been cleaned,