Hazards and Precautions| xi
Hazards and Precautions
Hazards
The following hazards are provided to help avoid injury:
Warning! Power Rating.
The instrument’s power supply or power cord must
be connected to a power receptacle that provides voltage and current within the
specified rating for the system. Use of an incompatible power receptacle may
produce electrical shock and fire hazards.
Warning! Electrical Grounding.
Never use a two-prong plug adapter to
connect primary power to the external power supply. Use of a two-prong
adapter disconnects the utility ground, creating a severe shock hazard. Always
connect the power cord directly to an appropriate receptacle with a functional
ground.
Warning! Service.
Only qualified technical personnel should perform service
procedures on internal components.
Warning! Accessories.
Only accessories which meet the manufacturer’s
specifications shall be used with the instrument.
Warning! Lubricants.
Do not apply lubricants to the microplate carrier or
carrier track. Lubricant on the carrier mechanism will attract dust and other
particles, which may obstruct the carrier path and cause the instrument to
produce an error.
Warning! Liquids.
Avoid spilling liquids on the instrument; fluid seepage into
internal components creates a potential for shock hazard or instrument
damage. If a spill occurs while a program is running, abort the program and
turn the instrument off. Wipe up all spills immediately. Do not operate the
instrument if internal components have been exposed to fluid.
Warning! Unspecified Use.
Failure to operate this equipment according to
the guidelines and safeguards specified in this manual could result in a
hazardous condition.
Warning! Direct Drain Waste.
If installed, the direct drain waste system
pumps waste fluids from the washer directly into a sink or tank, and, potentially
into public waste water systems. Because the waste may be a biohazard, you
must ensure that you are in compliance with your local or national
government’s laws regarding safe disposal of the waste.
Warning! Ultrasonic Energy.
Ultrasonic energy is present in the ultrasonic
cleaner reservoir (if equipped) when AUTOCLEAN/Quick Clean programs are
running. Avoid putting your fingers in the bath. Ultrasonic energy can be
destructive to human tissue.
Warning! Software Quality Control.
The operator must follow the
BioTek Instruments, Inc.
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......