188 | Chapter 7: Troubleshooting
Problem
Possible
Cause
What To Do
Y-axis.
Uneven
performance.
Plate carrier
leveling feet
damaged.
Replace plate carrier: contact BioTek TAC for assistance.
Note: "Q" model washers with Verify™ Technology must
recalibrate the Verify test plate after installing the new
carrier.
Washer Manifold Movement
Problem
Possible
Cause
What To Do
Manifold
position
error.
Manifold
movement is
blocked.
Check orientation of microplate; A1 should be in the left rear
corner of the plate carrier as you face the instrument.
Check for and remove any obstructions.
Ensure the manifold is installed properly.
Incorrect
manifold
selected.
Make sure the washer manifold setting matches the installed
manifold (96-, 192--tube).
LHC: Tools>Instrument Utilities>Washer>Manifold
Selection.
Keypad: Setup Menu>
g
>MANIF
Communication Errors
Here are some guidelines for troubleshooting communication errors between the
405 LS and the computer.
6045
: A potentially common error,
especially when using the Predefined
Protocols, a "serial write" error, is easily
fixed by correcting the
COM port setting
.
BioStack
communication error: when controlling the BioStack with the LHC, make
sure the onboard configuration setting is
Manual
.
810D
: Similarly, the 810D message appears when the instrument is busy, for
example when
AutoPrime
is running. The LHC can only talk to the instrument
when its main menu is displayed. Press the
Stop
button on the keypad, if desired,
to end the current process and return to the main menu.
405™ Microplate Washer LS
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......