Daily Maintenance | 121
Daily Maintenance
Daily maintenance involves flushing the washer with an appropriate reagent or
deionized water throughout the day. Routine rinsing helps to prevent the aspirate
and dispense tubes from clogging between runs. Flushing the devices with
deionized water is recommended at the end of the day for most applications.
The recommended
rinsing frequency
depends on the solutions currently in use:
l
When a solution containing surfactant is used throughout the day, perform
the rinsing procedure when the device is idle for more than 3 hours.
l
When the solution does
not
contain surfactant, consider rinsing at least once
an hour.
Run this protocol and enable
AutoPrime
to satisfy the daily maintenance
requirements:
l
DAY_RINSE
Make sure the supply bottles contain sufficient rinse solution and that the waste
bottles are empty before running the protocols.
Also see the additional maintenance procedures required when dispensing protein
solutions: Removing Protein Residuals and Fungi Growth on page 124.
AutoPrime
AutoPrime
automatically conditions the dispense tubes, priming them with the
specified volume, after a user-specified amount of idle time.
See AutoPrime on
next page
.
l
Press the
STOP
button to interrupt the AutoPrime routine when it is
underway.
l
Any interaction with the instrument via the keypad or the LHC resets the
interval clock.
l
AutoPrime only runs when the main menu, quick menu, or run completion
message is displayed on the keypad.
Overnight/Multi-Day Maintenance
Overnight/multi-day maintenance involves flushing all solutions out of the
instrument, and then periodically rinsing and soaking the tubes to keep them
moist. Here are three recommendations for accomplishing the task. Employ the
method that best suits your work flow:
BioTek Instruments, Inc.
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......