68 | Chapter 3: Operation
l
In a Wash step, select the
Show Wash Options
link and fill the Pre-
dispense checkbox.
Washer Flow Rates
The default setting for flow rates is 7, in the middle of the range for standard
tubing rates 3-11, 3 is the slowest, 11 is the fastest. Slower rates are recommended
for viscous fluids.
Dual 96-Tube Manifold
: When this manifold is installed, you can select rate 1 or 2
to use the low-flow tubing to dispense fluid slowly and gently to avoid damaging
well contents.
Rate
µL/tube/second
96 tube
192 tube
Cell Wash
1
116*
NA*
Cell Wash
2
134*
NA*
3
204
102
4
244
122
5
273
136.5
6
306
153
Default
7
325
162.5
8
352
176
9
375
187.5
10
391
195.5
11
418
209
*Cell Wash flow rates direct fluid through special low-flow tubing, which is not
compatible with the single 96-tube and 192-tube wash manifolds.
Aspirate Travel Rates
Rate
mm/sec
1
4.1 +1
Slowest
1 CW
4.1
2
5.0 +1
2 CW
5.0
3
7.3 +1
Default rate
3 CW
7.3