182 | Chapter 7: Troubleshooting
General Operation Troubleshooting
This is a list of potential performance problems and their solutions.
Startup
Problem
Possible Cause
What To Do
Display (Touch
screen) not on.
Power cord not
plugged in.
Check power connection.
Carrier/manifold
position error.
Manifold or
carrier is being
obstructed.
Remove obstruction.
Motor, sensor, or
electrical
problem.
Turn instrument off, wait at least 15 seconds, and
turn it back on. If the self-test does not pass, contact
BioTek TAC.
Misaligned
carrier or
manifold.
Contact BioTek TAC.
Incorrect
manifold setting.
Make sure the manifold setting matches the installed
manifold.
LHC: Tools>Instrument Utilities>Washer
Touch screen: Instrument>Options
Microplate Scratches
Problem
Possible Cause
What To Do
Scratches
on
microplate
bottom.
Microplate dispense or
aspiration height
adjustment too low.
Change the Dispense or Aspirate height, the Z-axis
position.
Microplate not properly
seated or strips not
level.
Reseat microplate in carrier or strips in holder.
Make sure the carrier is clean.
Try a different microplate or strip holder. If the
problem is unresolved, the carrier may have to be
realigned. Contact BioTek TAC.
405™ Microplate Washer LS
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......