Configure Verify™ Technology Utilities | 79
For example, if the actual sigma is 7.5, and mean volume is 250, then actual CV =
7.5/250 = 3%.
And:
CV
to
to
%
=
= 3 %
5 %
Verify
7.5 + 3
250 + 14
7.5 + 9
250 − 14
2
2
2
2
Configure Verify™ Technology Utilities
Tools>Instrument Utilities> Verify Manifold>Configure Utilities link
l
Raw Data folder
: This field shows the path and folder that LHC creates on your PC and
uses to store the test data. Every test creates an .xml data file that displays the test
results in the LHC. Regularly archiving these "Raw Data" files will ensure optimal viewing
performance.
l
Reports folder
: This customizable field shows where the text (.txt) files of the test
results are stored. To change the folder, click the link to browse to a desired location or
edit the path shown in the field.
Managing Test Results
l
Improve online viewing of test results
: If test results do not appear instantly after
a test, there may be too many raw data files in the
Raw Data folder
. Archive the data
files, i.e. save the files in a different folder to restore normal performance.
l
Archive files
: Develop a schedule that compliments your current QC/QA practices to
regularly archive Verify test raw data files. Use Windows Explorer to move data files
(.xml) and report files (.txt) to archive folders.
Test Plate Alignment
Procedure for calibrating the level sensor to work with a replacement test plate.
Depending on frequency of use, the Verify Test Plate will occasionally need to be
replaced due to wear or damage. When you receive a replacement plate from
BioTek, perform this procedure to update the washer's calibration data to match the
new plate.
See Replacement Procedure for Verify Test Plate on page 134
.
Sensor Calibration
This procedure serves two purposes:
l
Testing/determining the Slope and Offset volume coefficients for a specific Verify Test
Plate, i.e. a step in the procedure to replace a test plate.
l
Qualify the Verify™ Technology Sensor on page 167 to ensure it is performing to
specification.
BioTek Instruments, Inc.
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......