82 | Chapter 3: Operation
LHC:
Keypad:
1. Press
Setup Menu
.
2. Select
BIOSTK
.
3. Select
CONF
.
4. Select
BIOSTACK
.
Fill the BioStack checkbox in the main view
to enable the BioStack action buttons and
use them to design a protocol that delivers
and retrieves plates.
Important: When using LHC to control the 405
LS and the BioStack, the instrument's BioStack
configuration setting must be set to
MANUAL
,
not "BIOSTACK."
n
To Restack or not? Yes: to keep plates in the same order. No: to save time
when the plate sequence is unimportant.
5.
Verify
the setup: perform a protocol with 1 or 2 plates.
At the start of the day, power up the BioStack first, and then the 405 LS. BIOSTACK2WR:
Lift the BioStack’s gripper before turning it on.
Robotics integrators: CAD drawings of the physical dimensions of the 405 LS are
available upon request. Contact BioTek customer service.
Technical Note
: Only one of the two communication ports (COM port) on the
instrument can be used at a time. They cannot be used simultaneously. You
can use USB to connect the 405 LS to the computer or the RS232 serial port to
connect to a BioStack or similar robotic device. But you cannot use both ports
simultaneously, i.e. make sure only one cable is plugged in at a time.
Keypad Control
: When the BioStack is connected to your 405 LS, you are
controlling both instruments using the keypad. Before connecting the 405 LS to
your computer to download basecode or for other reasons, you must first
disconnect the BioStack from the 405 LS and change the Instrument Setting for the
BioStack: Press
Setup Menu> BIOSTK> CONF>MANUAL
.
BioStack Alignment Utility
Follow the detailed procedure in the BioStack Operator's Manual to use this
alignment utility to precisely align the BioStack's gripper with the microplate
405™ Microplate Washer LS
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......