Hazards and Precautions| xiii
and other hardware.
Caution: Chemical Compatibility.
Some chemicals may cause irreparable
damage to the instrument. The following chemicals have been deemed safe for
use in the instrument: buffer solutions (such as PBS), saline, surfactants,
deionized water, 70% ethyl, isopropyl, or methyl alcohol, 40% formaldehyde,
and 20% sodium hydroxide. Never use acetic acid, DMSO, or other organic
solvents. These chemicals may cause severe damage to the instrument.
Contact BioTek for more information and prior to using other questionable
chemicals.
Caution: Bovine Serum Albumin.
Solutions containing proteins, such as
bovine serum albumin (BSA), will compromise the instrument’s performance
over time unless a strict maintenance protocol is adhered to. See
Maintenance
procedures regarding BSA.
Caution: Power Supply.
Only use the power supply shipped with the
instrument. Operate this power supply within the range of line voltages listed
on it.
Caution: Disposal.
This instrument contains printed circuit boards and wiring
with lead solder. Dispose of the instrument according to Directive 2002/96/EC,
“on waste electrical and electronic equipment (WEEE),” or local ordinances.
Caution: Warranty.
Failure to follow preventive maintenance protocols may
void the warranty.
Caution: Shipping Hardware.
All shipping hardware (e.g., shipping bracket
etc.) must be removed before operating the instrument and reinstalled before
repackaging the instrument for shipment.
Caution: High Flow Pump Installation.
DO NOT plug the High Flow vacuum
pump cable into a wall outlet! Use the adapter provided with the pump to
connect the pump to the accessory outlet on the back of the washer. See the
Installation
instructions
.
Caution: Waste Sensor Port on 405 LS.
(For customers who have
purchased the BioStack Microplate Stacker.) Although the waste sensor port on
the back of the 405 LS is the same type as the 24-VDC power connector on the
back of the BioStack, if an external 24-VDC power supply is plugged into the
405 LS’s port,
it will permanently damage internal components
.
Caution: Electromagnetic Environment.
Per IEC 61326-2-6 it is the user’s
responsibility to ensure that a compatible electromagnetic environment for this
instrument is provided and maintained in order that the device will perform as
intended.
BioTek Instruments, Inc.
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......