92 | Chapter 3: Operation
Aspirate Settings:
Adjust the Y Position to align the
aspirate tubes near the well walls, if
available. Increase the Aspirate Height (Z
axis), leaving more residual volume in
the wells.
384-well Flat Magnet PN 7103017
The magnetic force (approx. 4300 Gauss)
is distributed in a vertical pattern,
column-wise, across the plate.
Adjust the X Position to align the
aspirate tubes away from the center of
the well, near the well walls.
Ring Magnets
Use this information about the VP magnets to fine-tune wash protocol settings:
96 Ring Magnet
7102216
384 Ring Magnet
7102215
96 Ring Magnet
PN: 7102216
This magnetic bead separator uses
329 of VP’s 52 MGO magnets (7094
Gauss). The magnets are arranged
around each well, pulling the
magnetic beads to the bottoms and
edges of the wells. Aspirate from
the center of the well (the default
position), when using this magnet.
384 Ring Magnet
PN: 7102215
This magnetic bead separator uses
425 of VP’s 52 MGO magnets (6994
Gauss). The magnets are aligned
with the intersections of the wells,
pulling the magnetic beads to the
Aspirate from the center of the well
(the default position), when using this
magnet.