194 | Chapter A: Error Codes
Code
Message
What to do
C03
Configuration parameter out of
range
Service Only.
Contact BioTek TAC.
1001
Bootcode powerup checksum
test failed
Contact BioTek TAC.
1002
Unknown error in bootcode
Contact BioTek TAC.
1003
Bootcode page program error
Contact BioTek TAC.
1004
Bootcode block size error (not
256)
Contact BioTek TAC.
1005
Invalid processor signature (not
1280,1281,2560,2561)
Contact BioTek TAC.
1006
Bootcode memory exceeded
Contact BioTek TAC.
1007
Invalid slave port
Contact BioTek TAC.
1008
Invalid response from slave
Contact BioTek TAC.
1009
Invalid processor detected
Contact BioTek TAC.
1010
Checksum error downloading
basecode
Contact BioTek TAC.
1250
Internal RAM test error on the
UI processor
Contact BioTek TAC.
1251
Internal RAM test error on the
MC processor
Contact BioTek TAC.
1260
Stack test error on the UI
Contact BioTek TAC.
1261
Stack test error on the MC
Contact BioTek TAC.
1400
No vacuum pressure detected
after turning on the vacuum
pump
Make sure the vacuum pump is turned on,
and not leaking; the waste bottle has not
overflowed and its caps are seated
correctly. If the pump is not running, the
fuse may be blown. But, before replacing the
fuse, first try to determine and remedy the
cause of the failure. If the pump sounds
strained or runs slowly, it may be damaged.
Review other related Troubleshooting
suggestions.
405™ Microplate Washer LS
Summary of Contents for 405 LS
Page 1: ...Operator s Manual Microplate Washer 405 LS ...
Page 2: ......
Page 56: ...36 Chapter 2 Installation 405 Microplate Washer LS ...
Page 134: ...114 Chapter 3 Operation 405 Microplate Washer LS ...
Page 190: ...170 Chapter 5 Qualification 405 Microplate Washer LS ...
Page 210: ...190 Chapter 7 Troubleshooting 405 Microplate Washer LS ...
Page 238: ......