Vacuum Filtration for Filter Plate Assays | 99
Set up the Vacuum Filtration module
Connect the Intermediate Waste System
Intermediate Waste System Setup
For this step you will need the
Vacuum Filtration
tubing, intermediate waste
bottle, and the control module. Review the tubing map on the front of the
control module. It is a good guide for this step.
Three lengths of tubing are supplied to set up the control module. Connect the
tubing to the control module and to the intermediate waste bottle cap:
Tubing
Control
Module
Waste Bottle Cap
1
Dual tubing connected by Luer
fitting, approx. 15”/38.1 cm
Intermediate
Waste Out
Smallest nipple with long
interior tube
2
Long single tube, approx. 15”/
38.1 cm
Intermediate
Vacuum Supply
Nipple without a tube inside
the bottle
BioTek Instruments, Inc.
Summary of Contents for 405 LS