104 | Chapter 3: Operation
Control the Vacuum Level
Vent plugs stored on side of carrier
Change the vent plug to control the vacuum level
The vacuum filtration plate carrier has a vent and ships with four vent plugs to
vary the vacuum levels:
Plug
Vent Diameter
Vacuum Level
mmHg kPa
No plug
0.067” (1.70 mm)
Lowest
-32
-4.3
Beige with hole
0.047” (1.19 mm)
Low
-118
-15.7
Gray
0.032” (0.81 mm)
Medium
-187
-24.9
Black
0.020 (0.51 mm)
High
-382
-50.9
Beige (solid)
0.00
Highest
-567
-75.5
Leave the vent open for the least amount of vacuum. Insert one of the plugs to
increase vacuum.
n
Vacuum pressure is affected by several factors like relative humidity,
barometric pressure, and mechanical tolerances, as well as, filter plate pore
size