Replacement Procedure for Verify Test Plate | 137
Volume coefficients - Slope and Offset
BioTek's Verify™ Technology level sensor has been calibrated at the factory to
work with a specific Verify Test Plate. These volume coefficients have been
gravimetrically determined and set for your washer-test plate combination.
Important:
Do not change these values unless specifically instructed to do so by BioTek,
as when replacing the test plate, for example.
Note: default values are Slope = 0.6 and Offset = -200.0.
BioTek Instruments, Inc.