TROUBLESHOOT
THE
A51
GSP
and
A52
L
CD
Use
this
pro cedure
when
the
LCD(Liquid
Crystal
Display)
is
unacceptable,
or
not
b eing
brigh
t.
1.
Run
the
Internal
T
est
3:
A51
GSP
.
The
A51
GSP
can
b e
c
hec
k
ed
using
the
in
ternal
test
3:
A51
GSP
,
if
the
test
fails,
the
4
Ch
1
5
and
4
Ch
2
5
LEDs
blink
sev
eral
time
and
a
few
b eeps
sound
at
the
end
of
the
test.
Then
the
analyzer
returns
the
con
trol
settings
to
the
p o
w
er-on
default
setting
v
alues.
a.
Press
4
PRESET
5,
4
SYSTEM
5,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
MENU
,
NNNNNNNNNNNNNNNNN
TESTS
,
4
3
5,
4
x1
5,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXECUTE
TEST
to
run
the
in
ternal
test
3.
When
this
test
starts,
4
Ch
1
5
LED
and
4
Ch
2
5
LED
are
turned
o.
b.
Chec
k
the
4
Ch
1
5
and
4
Ch
2
5
LEDs
and
the
b eeps
at
the
end
of
the
test.
If
no
b eep
sound
and
the
LEDs
don't
blink,
the
A51
GSP
is
probably
w
orking.
Con
tinue
with
the
next
Che
ck
the
Two
LEDs
on
A51
GSP.
If
a
b eep
sounds
and
the
LEDs
blink
one
time,
the
A51
GSP
c
hip
is
faulty
.
Replace
the
A51
GSP
.
If
t
w
o
b eep
sound
and
the
LED
blinks
t
w
o
time,
the
A51
GSP's
DRAM
is
faulty
.
Replace
the
A51
GSP
.
If
three
b eep
sound
and
the
LED
blinks
three
time,
the
A51
GSP's
VRAM
is
fault
y
.
Replace
the
A51
GSP
.
2.
Check
the
A52
LCD(Liquid
Crystal
Display)
The
A52
LCD
can
b e
tested
using
the
in
ternal
test
70
to
74.
a.
Press
4
PRESET
5,
4
SYSTEM
5,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SERVICE
MENU
,
NNNNNNNNNNNNNNNNN
TESTS
,
4
7
5,
4
0
5,
4
x1
5,
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
EXECUTE
TEST
NNNNNNNNNNNNNNNNNNNNNNNNNN
CONTINUE
to
run
the
in
ternal
test
70,
and
run
the
other
tests
with
the
same
manner.
b.
If
an
y
defects
on
the
LCD,
replace
the
LCD.
c.
If
no
correct
patterns
are
display
ed,
c
hec
k
the
A54
In
v
erter.
8-12
Digital
Control
Troubleshooting
Summary of Contents for 4395A
Page 10: ......
Page 26: ......
Page 34: ......
Page 77: ...Figure 2 17 B R Magnitude Ratio Phase Dynamic Accuracy Test Setup 2 Performance Tests 2 43 ...
Page 167: ...Figure 5 1 Adjustment Hardware Setup Adjustments 5 5 ...
Page 186: ...Figure 5 13 Receiver Gain Adjustment Location 5 24 Adjustments ...
Page 190: ...Figure 5 16 Receiver Flatness Adjustment Setup 1 MHz 5 28 Adjustments ...
Page 194: ...Figure 5 20 DC Bias Adjustment Setup 2 5 32 Adjustments ...
Page 196: ...Figure 6 1 Troubleshooting Organization 6 2 Troubleshooting ...
Page 206: ...Figure 7 1 Power Supply Lines Simplified Block Diagram 7 2 Power Supply Troubleshooting ...
Page 212: ...Figure 7 5 A1 CPU Connector Locations 7 8 Power Supply Troubleshooting ...
Page 220: ...Figure 8 1 Digital Control Group Simplified Block Diagram 8 2 Digital Control Troubleshooting ...
Page 240: ...Figure 10 1 Top View Major Assemblies 10 4 Replaceable Parts ...
Page 292: ...Table A 2 Manual Changes by Firmware Version Version Make Manual Changes A 2 Manual Changes ...
Page 308: ......
Page 311: ...Figure B 1 Power Cable Supplied Power Requirement B 3 ...
Page 312: ......
Page 342: ......