![Agilent Technologies 4284A Operation Manual Download Page 299](http://html.mh-extra.com/html/agilent-technologies/4284a/4284a_operation-manual_2867874299.webp)
COrrection
Subsystem
:SPOT
<n > :FREQuency
The
:SPOT <n >:FREQuency
command
sets
the
frequency
p oints
(FREQ1,
FREQ2
or
FREQ3)
for
the
sp ecied
frequency
p oint
correction.
The
:SPOT <n >:FREQuency?
query
returns
the
curren
t
settings
of
the
frequency
p oints
(
FREQ1,
FREQ2
or
FREQ3).
Command
Syntax
CORRection:SPOT <n >:FREQuency
<value>
Where,
<value
>
is
the
NR1,
NR2,
or
NR3
format
<n
>
is:
1
F
requency
setting
for
FREQ1
p oint
2
F
requency
setting
for
FREQ2
p oint
3
F
requency
setting
for
FREQ3
p oint
Example
OUTPUT
717;"CORR:SPOT1:FREQ
2KHZ"!
Set
2
kHz
to
FREQ1
Note
A
sux
m
ultiplier
and
a
sux
unit,
HZ
(hertz),
can
b e
used
with
this
command.
Either
MAHZ
and
MHZ
can
b e
used
as
the
sux
m
ultiplier
for
MHz
(1E6
Hz).
Query
Syntax
CORRection:SPOT <n >:FREQuency?
Query
Response
Returned
format
is
:
<NR3><NL^END>
Example
10
OUTPUT
717;"CORR:SPOT1:FREQ?"
20
ENTER
717;A
30
A
40
END
Note
If
this
query
is
receiv
ed
when
the
List
Sw
eep
parameter
is
set
to
an
ything
other
than
frequency
,
error
0230
Data
corrupt
or
stale
will
o ccur.
Command
Reference
8-59
Summary of Contents for 4284A
Page 12: ......
Page 37: ...Figure 1 1 Power Cable Supplied Installation and Set Up Guide 1 3 ...
Page 54: ...Figure 2 5 Display Pages 1 3 2 12 Overview ...
Page 55: ...Figure 2 5 Display Pages 2 3 Overview 2 13 ...
Page 56: ...Figure 2 5 Display Pages 3 3 2 14 Overview ...
Page 58: ...Figure 2 7 Softkey Selection Example 2 16 Overview ...
Page 60: ...Figure 3 1 Available Fields on the MEAS DISPLAY Page 3 2 DISPLAY FORMAT Menu ...
Page 61: ...Figure 3 2 Available Softkeys on the MEAS DISPLAY Page DISPLAY FORMAT Menu 3 3 ...
Page 66: ...Figure 3 4 Effective Measuring Range Oscillator Level 2V or 20 mA 3 8 DISPLAY FORMAT Menu ...
Page 91: ...Figure 3 15 Available Fields on the LIST SWEEP DISPLAY Page DISPLAY FORMAT Menu 3 33 ...
Page 96: ......
Page 99: ...Figure 4 2 Available Softkeys on the MEAS SETUP Page MEAS SETUP Menu 4 3 ...
Page 104: ...Figure 4 5 Available Operating Area for the ALC Function 4 8 MEAS SETUP Menu ...
Page 113: ...4 Press NNNNNNNNNNN YES to reset the 4284A MEAS SETUP Menu 4 17 ...
Page 115: ...Figure 4 8 Available Fields on the CORRECTION Page MEAS SETUP Menu 4 19 ...
Page 116: ...Figure 4 9 Available Softkeys on the CORRECTION Page 4 20 MEAS SETUP Menu ...
Page 131: ...Figure 4 15 Available Fields on the LIMIT TABLE SETUP Page MEAS SETUP Menu 4 35 ...
Page 145: ...Figure 4 21 Available Fields on the LIST SWEEP SETUP Page MEAS SETUP Menu 4 49 ...
Page 152: ......
Page 158: ...Figure 5 4 Available Fields on the SYSTEM CONFIG Page 5 6 Catalog System Configuration ...
Page 228: ...Figure 7 19 Standard Event Status Register 7 28 Remote Control ...
Page 240: ......
Page 361: ...Caution The memory card should be removed before packing the 4284A General Information 9 29 ...
Page 362: ......
Page 432: ......
Page 436: ......
Page 440: ......
Page 460: ......