![Agilent Technologies 4284A Operation Manual Download Page 274](http://html.mh-extra.com/html/agilent-technologies/4284a/4284a_operation-manual_2867874274.webp)
LIST
Subsystem
:CURRent
The
:CURRent
command
clears
the
previous
list
sw
eep
p oint
table,
and
sets
the
oscillator
curren
t
lev
el
sw
eep
p oints.
The
:CURRent?
query
returns
the
curren
t
settings
of
curren
t
sw
eep
p oints.
Command
Syntax
LIST:CURRent
<value>
2
, <value>*
3
*Max.
10
sw
eep
p oints
Where,
<value>
is
the
NR1,
NR2,
or
NR3
format
Example
OUTPUT
717;"LIST:CURR
100MA"
!Set
100mA
to
point
1
OUTPUT
717;"LIST:CURR
1E-2,2E-2,3E-2,4E-2"
!
Set
10
mA
to
point
1,
...
40
mA
to
point
4
Note
A
sux
m
ultiplier
and
a
sux
unit,
A
(amp
ere),
can
b e
used
with
this
command.
Query
Syntax
LIST:CURRent?
Query
Response
Returned
format
is
:
<NR3>
2
,<NR3> *
3
<NL^END>
Example
10
DIM
A$[100]
20
OUTPUT
717;"LIST:CURR?"
30
ENTER
717;A$
40
A$
50
END
Note
If
this
query
is
receiv
ed
when
the
List
Sw
eep
parameter
is
set
to
an
ything
other
than
curren
t,
error
0230
Data
corrupt
or
stale
will
o ccur.
8-34
Command
Reference
Summary of Contents for 4284A
Page 12: ......
Page 37: ...Figure 1 1 Power Cable Supplied Installation and Set Up Guide 1 3 ...
Page 54: ...Figure 2 5 Display Pages 1 3 2 12 Overview ...
Page 55: ...Figure 2 5 Display Pages 2 3 Overview 2 13 ...
Page 56: ...Figure 2 5 Display Pages 3 3 2 14 Overview ...
Page 58: ...Figure 2 7 Softkey Selection Example 2 16 Overview ...
Page 60: ...Figure 3 1 Available Fields on the MEAS DISPLAY Page 3 2 DISPLAY FORMAT Menu ...
Page 61: ...Figure 3 2 Available Softkeys on the MEAS DISPLAY Page DISPLAY FORMAT Menu 3 3 ...
Page 66: ...Figure 3 4 Effective Measuring Range Oscillator Level 2V or 20 mA 3 8 DISPLAY FORMAT Menu ...
Page 91: ...Figure 3 15 Available Fields on the LIST SWEEP DISPLAY Page DISPLAY FORMAT Menu 3 33 ...
Page 96: ......
Page 99: ...Figure 4 2 Available Softkeys on the MEAS SETUP Page MEAS SETUP Menu 4 3 ...
Page 104: ...Figure 4 5 Available Operating Area for the ALC Function 4 8 MEAS SETUP Menu ...
Page 113: ...4 Press NNNNNNNNNNN YES to reset the 4284A MEAS SETUP Menu 4 17 ...
Page 115: ...Figure 4 8 Available Fields on the CORRECTION Page MEAS SETUP Menu 4 19 ...
Page 116: ...Figure 4 9 Available Softkeys on the CORRECTION Page 4 20 MEAS SETUP Menu ...
Page 131: ...Figure 4 15 Available Fields on the LIMIT TABLE SETUP Page MEAS SETUP Menu 4 35 ...
Page 145: ...Figure 4 21 Available Fields on the LIST SWEEP SETUP Page MEAS SETUP Menu 4 49 ...
Page 152: ......
Page 158: ...Figure 5 4 Available Fields on the SYSTEM CONFIG Page 5 6 Catalog System Configuration ...
Page 228: ...Figure 7 19 Standard Event Status Register 7 28 Remote Control ...
Page 240: ......
Page 361: ...Caution The memory card should be removed before packing the 4284A General Information 9 29 ...
Page 362: ......
Page 432: ......
Page 436: ......
Page 440: ......
Page 460: ......