Hardware Troubleshooting
9 - 4
Modified: 9/17/08
MedWeld 4000 Operator’s Guide
M-032116 Programs #T03300, #T02300 and #T02400
Control and DEP-
100S cannot
communicate
(DEP and MedLAN
LEDs may be off or
blinking)
Bad connection between
weld processor and DEP-
100S.
Check connection. Verify that it is
secure.
Check for bad/faulty connector and/or
cable.
Cycle power.
If networking, no network
power supply or no external
120 VAC source connected
to the power supply.
Make sure you have a network power
supply and that external 120 VAC is
connected.
Faulty DEP-100S, WTC
processor or Network
Power Pack.
Replace as necessary.
Incorrect network address
for weld control.
Verify the welder ID and network
address of the device.
Communication over
the MedLAN channel
is inoperable.
(MedLAN LEDs may
be off or blinking)
Improper or loose MedLAN
wiring.
Check for secure and correct wiring.
Two or more controls have
the same network address
number.
Check MedLAN address of each
control. Each control must have a
UNIQUE address.
Faulty WTC processor,
faulty DEP-100S or faulty
Network Power Pack.
Replace as necessary.
Symptom
Probable Cause
Corrective Action
Содержание MedWeld 4000
Страница 77: ...9 cdQ Y W dXU UTGU T _TYVYUT UTGU T UbQd_bµc 7eYTU b_WbQ c D D Q T D This page is intentionally left blank ...
Страница 153: ...Weld Schedules 4 54 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 255: ...Hardware Troubleshooting 9 10 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 257: ...Application Note Current Compensation 2 ML 010041 Application Note Current Compensation ...
Страница 261: ...Application Note Cascade Welding 4 ML 010041 Application Note Cascade Firing ...