Setup Parameters
MedWeld 4000 Operator’s Guide
Modified: 9/17/08
5 - 7
M-032116 Programs #T03300, #T02300 and #T02400
I/O UPDATE TIMEOUT: (
T03300 only
)(FAULT)
This fault may occur in one of these conditions:
•
The processor entered Program Mode while the control was
executing a weld sequence.
•
The WTC weld processor loses communications for more than
the time specified in the I/O Update Timeout (mSEC): nnn
setup parameter. The fault may be set at any time, including
when the control is initiating a schedule. If this occurs, the
schedule is aborted.
When the control is idled and not initiated, the timeout interval
is fixed at 10 sec.
•
When the weld processor cannot communicate over the rack
backplane. If this timeout occurs, major system bus faults are
suspect.
•
If the weld processor moves out of Run Mode while the control
is executing a schedule.
RIO COMM ERROR/NOT IN RUN MODE:
(
T03300 only
) (FAULT)
This fault occurs
•
If the host connected to the RIO link moves out of Run Mode,
or
•
The link encounters a fatal communication error while the
control is executing a schedule.
In either case, this fault is flagged regardless of the state of the I/O
VALID input.
WELD PROCEED:
(FAULT) (ALERT)
This fault is generated if the WELD PROCEED input does not
become active within the number of cycles specified in the function
WAIT nnn CY. FOR WELD PROCEED.
PRESSURE SWITCH:
(FAULT) (ALERT
This fault is generated if the PRESSURE SWITCH input does not
become active within the number of cycles specified in Function
#68 (WAIT nnn CY. FOR PRESSURE SWITCH) or if the weld
initiate is removed while the control is waiting in Function #69
(WAIT FOR PRESSURE SWITCH).
Содержание MedWeld 4000
Страница 77: ...9 cdQ Y W dXU UTGU T _TYVYUT UTGU T UbQd_bµc 7eYTU b_WbQ c D D Q T D This page is intentionally left blank ...
Страница 153: ...Weld Schedules 4 54 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 255: ...Hardware Troubleshooting 9 10 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 257: ...Application Note Current Compensation 2 ML 010041 Application Note Current Compensation ...
Страница 261: ...Application Note Cascade Welding 4 ML 010041 Application Note Cascade Firing ...