Hardware Troubleshooting
MedWeld 4000 Operator’s Guide
Modified: 9/17/08
9 - 9
M-032116 Programs #T03300, #T02300 and #T02400
No line voltage
reading or no
zero-crossing
Main breaker switch is OFF.
Turn ON the breaker.
No 480/600 VAC is connected to the
top of the breaker.
Verify that supply voltage (480 or 600
VAC) is connected to the top of the
breaker. (On a single-pack control, supply
voltage is connected at L1 and L2.)
Faulty circuit breaker.
Supply voltage IS connected to the
breaker. Verify that power is available on
the load side of the circuit breaker.
Faulty cable between firing and
processor card.
Replace cable.
If “No Zero Crossing Sync” occurs
during a weld, there may be a loose
connection or weak welding bus.
(On the 1,200-A. inverter, the zero-
crossing signal comes through the
gate board on the SCR pack.)
Tighten loose connections. Install weld
bus capacitors. Re-distribute loads on the
weld bus. Install a stiffer power supply.
Faulty firing card.
Replace firing card.
Faulty weld processor.
Replace weld processor.
Problem
Possible Cause
Possible Solution
Содержание MedWeld 4000
Страница 77: ...9 cdQ Y W dXU UTGU T _TYVYUT UTGU T UbQd_bµc 7eYTU b_WbQ c D D Q T D This page is intentionally left blank ...
Страница 153: ...Weld Schedules 4 54 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 255: ...Hardware Troubleshooting 9 10 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 257: ...Application Note Current Compensation 2 ML 010041 Application Note Current Compensation ...
Страница 261: ...Application Note Cascade Welding 4 ML 010041 Application Note Cascade Firing ...