Fault Conditions
8 - 18
Modified: 9/17/08
MedWeld 4000 Operator’s Guide
M-032116 Proograms #T03300, #T02300 and #T02400
ISO CNTR
OFF
Fault
DEP-100S displays
INVERTER ISOLA-
TION CONTACTOR
OFF FAULT.
(No Inverter Charg-
ing Request/Inverter
Ready Request
signal.)
Control transformer is not
supplying voltage/current.
Replace transformer.
Failure of output module
on chassis.
Replace the output card.
Control transformer fuse
failure.
Replace fuses.
Inverter input failure.
Replace inverter as necessary.
CURRENT
REG
Fault
DEP-100S displays
CURRENT REGU-
LATION FAULT.
(The inverter cannot
reach the target cur-
rent.)
DEP-100S on-time is
always at maximum. Tar-
get heat is too high.
Increase mains voltage up to
480 VAC.
Reduce secondary impedance
(Possibilities: Run shunts to gun
installed closely together; Install
shorter shunts if possible; Install
larger shunts.)
Increase weld transformer
secondary voltage.
Target current reached as
inverter begins weld cycle.
DEP-100S shows average
current that is more than
requested current. (Small
secondary current
requested.)
Current does not decrease
from higher to lower weld
current because DC sec-
ondary current does not
decay quickly enough.
Decrease weld transformer
secondary voltage.
Add a 1-cycle wait between
welds;
or
Add a 1/2-cycle impulse
function between welds.
LED On
Inverter
Symptom
Probable Cause
Corrective Action
Содержание MedWeld 4000
Страница 77: ...9 cdQ Y W dXU UTGU T _TYVYUT UTGU T UbQd_bµc 7eYTU b_WbQ c D D Q T D This page is intentionally left blank ...
Страница 153: ...Weld Schedules 4 54 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 255: ...Hardware Troubleshooting 9 10 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 257: ...Application Note Current Compensation 2 ML 010041 Application Note Current Compensation ...
Страница 261: ...Application Note Cascade Welding 4 ML 010041 Application Note Cascade Firing ...