Advanced Software Features
MedWeld 4000 Operator’s Guide
Modified: 9/17/08
7 - 5
M-032116 Programs #T03300, #T02300 and #T02400
Target current
=
(%I programmed in Function #20 + Stepper boost) x C-Factor
=
(50 + 2) x 170 = 8,840 A.
To determine the High Current Limit:
High Current Limit
= (Target x %High Limit/100)
= 8,840 + (8,840 x 10/100)
= 8,840 + (884) = 9,724 A.
To determine the Low Current Limit:
Low Current Limit
= Target – (Target x %Low Limit/100)
= 8,840 – (8,840 x 20/100)
= 8,840 – (1,768) = 7,072 A.
ACC Firing Mode
In this firing mode, the target value is established by the following
calculation:
Target current
= A. in Function #30 + Stepper boost
An example of how these calculations are used is shown below.
•
Function #30 is WELD 08 CY 11000 A. SEC
•
Stepper Boost is 700 A. SEC
•
High Current Limit = 10%
•
Low Current Limit = 20%
To determine the Target Current:
Target current
= A. in Function #30 + stepper boost
= 11,000 + 700 = 11,700 A.
To determine the High Current limit:
High Current Limit
= (Target x %High limit/100)
= 11,700 + (11,700 x 10/100)
= 11,700 + (1170) = 12,870 A.
To determine the Low Current Limit:
Low Current Limit
= Target – (Target x %Low limit/100)
= 11,700 – (11,700 x 20/100)
= 11,700 – 2,340 = 9,360 A.
Содержание MedWeld 4000
Страница 77: ...9 cdQ Y W dXU UTGU T _TYVYUT UTGU T UbQd_bµc 7eYTU b_WbQ c D D Q T D This page is intentionally left blank ...
Страница 153: ...Weld Schedules 4 54 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 255: ...Hardware Troubleshooting 9 10 Modified 9 17 08 MedWeld 4000 Operator s Guide M 032116 Programs T03300 T02300 and T02400 ...
Страница 257: ...Application Note Current Compensation 2 ML 010041 Application Note Current Compensation ...
Страница 261: ...Application Note Cascade Welding 4 ML 010041 Application Note Cascade Firing ...