LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Parameter Bezel Replacement
312
8
To remove the parameter bezel from the rear case
:
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy PCB assembly as described in
Removing the System (A01)/Therapy (A04) PCB Assembly (p. 233)
3.
Remove the OEM PCB as described in
Removing the OEM PCB (A06) (p. 262)
4.
Disconnect the power/contact PCB (W05) from the power board at J12 (see
5.
Remove the power/system cable (W01) from the power board at J17.
6.
If present, disconnect the CO2 inlet tube.
7.
If present, carefully disconnect the CO2 cable connector (W28) from CO2 adapter cable (W30).
8.
If present, disconnect the NIBP tube at tube fitting.
9.
If present, carefully disconnect the SpO2 connector flex cable (W22) (
) from J1 of the SpO2 PCB.
CAUTION
POSSIBLE COMPONENT DAMAGE
The SpO2 connector flex cable can be damaged during disassembly. If SpO2 connector is
damaged, replace the SpO2 flex cable using the
MASIMO Sp02 Panel Mount Cable Repair Kit (MIN 3305431-026) (p. 514)
kit.
10. If present, remove the CO2 door.
11. Remove the parameter bezel label from the bezel.
12. Remove the two bezel mounting screws (
13. Lift out the parameter bezel (
) from the rear case.