LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Backlight/Interface PCB Cable (W06) Replacement
213
8
Backlight/Interface PCB Cable (W06) Replacement
Refer to
Inside Front Case Diagram (p. 189)
Removing the Backlight/interface PCB (W06) Cable
To remove the backlight/interface W06 cable (
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Press the connector locking tab and disconnect the W06 cable from the interface
PCB at J37.
3.
Place even pressure on the cable and ease the W06 cable out of the connector on
the Backlight PCB.
Installing the Backlight/interface PCB (W06) Cable
To install the backlight/interface W06 cable into the front case:
1.
2.
If interface PCB is present, connect the W06 cable to the interface PCB at J37.
3.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
P37
P49
to A05 - J37
to A08 - J49
Figure 8.22—Backlight PCB/interface PCB
cable connections