LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
CO2 Inlet Connector Cable (W28) Replacement
352
8
Installing the CO2 Inlet Connector Cable (W28)
To install the CO2 inlet connector cable:
NOTE:
When replacing the CO2 inlet connector cable assembly, use
CO2 Connector Repair Kit (MIN 3305431-014) (p. 506)
1.
Set the CO2 connector adapter (
) in place on the face of the new bezel.
2.
Place the new CO2 connector retainer (
) in place on the underside of the bezel and
tighten together with four new screws (
); torque to 2.0 in-lb using a P0 bit.
3.
Slide the CO2 inlet connector cable (W28) (
) into the new CO2 connector retainer.
4.
Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
5.
Install the system/therapy PCB assembly as described in
Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239)
6.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
CAUTION
POSSIBLE MOISTURE LEAKAGE
When installing the W28 cable, apply 2 in-lb of
torque when installing the screw retaining the W28 cable to help prevent ingress of
fluids.