LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
System (A01)/Therapy (A04) PCB Assembly Replacement
232
8
System (A01)/Therapy (A04) PCB Assembly Replacement
The system/therapy PCB assembly is removed as a unit and then separated. Refer to
System/Therapy PCB Assembly Diagrams and Parts List (p.
for diagrams of this assembly. See
for a connection diagram of the therapy PCB and
for a connection diagram of the system PCB.
Replacement consists of the following procedures:
•
Removing the System (A01)/Therapy (A04) PCB Assembly (p. 233)
•
Separating the System PCB (A01) (p. 244)
•
Replacing the System PCB (A01) (p. 246)
•
Separating the Therapy PCB (A04) (p. 247)
•
Replacing the Therapy PCB (A04) (p. 250)
•
Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239)