LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Interface PCB (A05) Replacement
194
8
Installing the Interface PCB (A05)
Refer to
Interface PCB Replacement Figures (p. 190)
Inside Front Case Diagram (p. 189)
.
To install the interface PCB into the front case:
NOTE:
If a new interface PCB is required, install the
Interface PCBA Repair Kit, V1 (MIN 3305431-005)
PCBA Repair Kit, V2 (MIN 3305431-029)
1.
Attach the interface PCB to the bracket with the eight new screws (
); torque to 6.8 in-lb (see
).
2.
Install the ground strap (W19) with one new screw (
); torque to 6.8 in-lb (see
3.
Install the connectors clockwise, as follows (make sure all connector locking tabs are positioned to secure their respective
connectors) (see
~ J31 – Install the main keypad interface PCB cable (W13,
).
~ J37 – Install the backlight interface PCB cable (W06,
~ J36 – Install the LCD interface PCB cable (W18,
~ J32 – Install the printer control keypad interface PCB cable (W12,
~ J34 – Install the speaker interface PCB wire (W17,
~ J35 – Install the printer interface cable (W16,
).
~ J33 – Install the speed dial cable (W15,
).
4.
If the interface PCB was replaced, install the system/interface cable to the interface PCB as described in
Cable (W04) Replacement (p. 212)
5.
Reassemble the case as described in
Reassembling the Case (p. 184)
.