LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Backlight PCB (A08) Replacement
197
8
Installing the Backlight PCB (A08)
To install the backlight PCB into the front case:
NOTE:
If a new backlight PCB is required,
install the
1.
backlight PCB.
2.
Connect the two end connectors to the LCD display
assembly.
3.
Connect the backlight/interface PCB cable (W06)
(
) to backlight power connector.
4.
Place the backlight inverter on the display assembly
and install with two new screws (
torque to 6.8 in-lb.
5.
Feed excess length of LCD wires into backlight
inverter cover.
6.
Install the interface PCB as described in
.
7.
Reassemble the case as described in
.
Figure 8.14—Screw location and cable connection for backlight PCB
screw locations
W06 - power connector
LCD connection
point
LCD connection
point