Preventive Maintenance
Setting/Resetting the Maintenance Prompt Interval
150
6
Setting/Resetting the Maintenance Prompt Interval
The MAINTENANCE DUE message can be set up to appear at selected intervals (3, 6, or 12 months). When this time interval is reached, the
message appears continuously for 10 minutes each time the device is turned on.
After completing a scheduled maintenance, reset the maintenance prompt interval timer to clear the MAINTENANCE DUE message and begin
the count for the next scheduled maintenance.
To change the scheduled maintenance interval:
1.
Enter Service mode as described in
.
2.
Select MAINT PROMPT to display the Service/Maint Prompt submenu, which shows the NEXT PROMPT date for scheduled
maintenance.
3.
Select INTERVAL.
4.
Select the desired interval. The factory default setting is OFF.
To clear the MAINTENANCE DUE message after scheduled maintenance is completed:
1.
Select RESET on the Service/Maint Prompt submenu. The NEXT PROMPT date is revised to the new scheduled maintenance
date.
2.
Turn the device OFF.