LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
NIBP (A21)/CO2 (A23) Module Replacement
293
8
2.
For CO2 option:
NOTE:
When replacing the CO2 module, use
Kit (MIN 3305431-002) (p. 497)
.
a.
) to the mounting bracket
with four new screws (
); torque to 4 in-lb.
b.
Connect the OEM PCB/CO2 module cable (W26) to the A23 CO2 PCB
at J4.
c.
Connect the CO2 adapter cable (W30) to the A23 CO2 PCB at J1.
To install the A21 NIBP, A23 CO2, and SpO2 PCBs:
(Continued) 12 steps, (Page 2 of 6)
CO2 option only
screws (4 places)
Figure 8.87—CO2 sub-assembly screw locations