LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Therapy Service Codes
134
5
a126
DE_ERROR_ADC_READ (Cannot read from ADC.)
a127
DE_ERROR_TEST_MODE (Must be idle to switch to test mode.)
a129
DE_ERROR_XFER_CABLE (Defib HW error.)
a12a
DE_ERROR_XFER_PADDLE (Defib HW error.)
a12c
DE_ERROR_CHG_INHIBIT (Defib HW error.)
a12d
DE_ERROR_CHG_ENABLE_FAIL (Charge enable feedback indicates not charging.)
a12e
DE_ERROR_BTE_FAULT (Cedar BTE Fault Line State.)
a12f
DE_ERROR_BTE_FAULT_CLEARED (Cedar BTE Fault Line State.)
a130
DE_ERROR_BTE_RESET (Defib HW error.)
a131
DE_ERROR_NO_BTE_HW (Biphasic HW not found.)
a132
DE_ERROR_NO_BTE_XFER (Defib HW error.)
a133
DE_ERROR_BTE_CEDAR_DRV_HI (A/D high bit sticky.)
a201
PA_ERROR_LONG_WDT (Long watchdog timer test failed.)
a202
PA_ERROR_SHORT_WDT (Short watchdog timer test failed.)
a203
PA_ERROR_SCI_RCV (SCI received overrun, framing or parity. Unit used near high EMI causing these issues.)
Table 5.13—Initial Character a, Therapy Service Codes (TH)
Code
Service Code Description
Corrective
Action Code