LIFEPAK 15 Monitor/Defibrillator
Service Manual
Troubleshooting
Troubleshooting Chart
103
5
12-Lead ECG Tests
Inappropriate screen message response Verify test setup and retry test.
Check ECG cable; replace if necessary.
Possible W07 ECG Connector Cable failure.
Possible A01 System PCB failure.
ECG gain out of tolerance
Verify test setup and retry test.
Check ECG cable; replace if necessary.
Possible A01 System PCB failure.
3-Lead ECG Tests
5-Wire ECG Tests
Inappropriate screen message response Verify test setup and retry test.
Check ECG cable; replace if necessary.
Possible W07 ECG Connector Cable failure.
Possible A01 System PCB failure.
ECG gain out of tolerance
Verify test setup and retry test.
Check ECG cable; replace if necessary.
Possible A01 System PCB failure.
QUIK-COMBO ECG
Tests
ECG gain out of tolerance
Verify test setup and retry test.
Check therapy cable; replace if necessary.
Possible A04 Therapy PCB failure.
Possible A01 System PCB failure.
ECG fast restore out of tolerance
Verify test setup and retry test.
Check therapy cable; replace if necessary.
Possible A04 Therapy PCB failure.
Table 5.1— Corrective Actions (Continued)
Area
Observed Symptom
Suggested Corrective Action