LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
USB Flex Module (W14) Replacement
339
8
Removing the W14 - USB Flex Module
To remove the USB flex module (W14) (
):
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Remove the system/therapy board as described in
Removing the System (A01)/Therapy (A04) PCB Assembly (p. 233)
.
3.
Remove the parameter bezel as described in
Removing the Parameter Bezel (p. 311)
.
4.
From inside the rear case, remove the screw (
) holding the spring clip (
Discard the screw.
5.
From the outside rear of the case, loosen the two retained screws and remove the USB access door (
6.
Remove the two USB connector mounting screws (
). Discard the screws.
7.
Remove the USB flex assembly.
Installing the W14 - USB Flex Module
To install the USB flex module (W14) (
):
1.
From inside of the rear case, place the USB flex module into slot in the rear case.
2.
); torque to 6.8 in-lb.
3.
From the outside of the rear case, install two screws (
) into the USB flex module; torque to 6.8 in-lb.
4.
Install the USB access door (
) by tightening the two retained screws; torque to 6.8 in-lb.
5.
Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
6.
Install the system/therapy PCB assembly as described in
Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239)
7.
Reassemble the case as described in
Reassembling the Case (p. 184)
.