LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Temperature Cable Assembly (W35) Replacement
358
8
Removing the Temperature Cable Assembly (W35)
To remove the temperature cable assembly from the rear case (refer to
Inside Rear Case Diagrams (p. 228)
1.
Disassemble the case as described in
Disassembling the Case (p. 181)
.
2.
Disconnect the temperature cable assembly (
) from the J7 connector on the system PCB (A01).
3.
Remove the parameter bezel as described in
Removing the Parameter Bezel (p. 311)
.
4.
Remove the four screws and captured washers (
) that secure the temperature connector to the parameter bezel.
5.
Remove the temperature connector and gasket. Discard the screws and gasket.
Installing the Temperature Cable Assembly (W35)
To install the temperature cable assembly into the rear case (refer to
Inside Rear Case Diagrams (p. 228)
NOTE:
When installing a new temperature cable assembly, use the
Temperature Cable Repair Kit (MIN 3305431-027) (p. 515)
1.
Place the temperature connector and new gasket (
) into the second connector mounting hole up from the bottom of
the parameter bezel. Secure the temperature connector with four new screws (
); torque to 6.8 in-lb. (See
2.
Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
3.
Connect the temperature cable assembly (
) to the system PCB (A01) at J7.
4.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
CAUTION
POSSIBLE MOISTURE LEAKAGE
When installing the cable, use new seal to help prevent ingress of fluids.