LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Temperature Cable Assembly (W35) Replacement
357
8
Temperature Cable Assembly (W35) Replacement
Temperature cable assembly replacement consists of the following procedures:
•
Removing the Temperature Cable Assembly (W35) (p. 358)
•
Installing the Temperature Cable Assembly (W35) (p. 358)
Figure 8.121—Temperature cable parts and screw locations
Temperature
cable
screw
locations
(4 places)
Connector
gasket
Note:
Align ECG
Ferrite as
shown prior
to
installation
into unit.