LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
ECG Connector Cable (W07) Replacement
330
8
Installing the ECG Connector Cable (W07)
Refer to
To install the ECG connector cable into the rear case (refer to
Inside Rear Case Diagrams (p. 228)
:
NOTE:
When installing a new ECG cable, use the
ECG Connector Repair Kit (MIN 3305431-010) (p. 503)
1.
Mount the ECG receptacle contact cable (
). Orient the ECG receptacle groove to the
notch in the seal.
2.
Capture the cable ferrite between the two cable support brackets (
) and line up the ECG receptacle.
3.
Secure the ECG connector cable and support brackets to the parameter bezel with the four new screws (
); torque to
6.8 in-lb.
4.
Install the parameter bezel as described in
Installing the Parameter Bezel (p. 313)
.
5.
) to the J6 connector on the system PCB (A01).
6.
Reassemble the case as described in
Reassembling the Case (p. 184)
.
CAUTION
POSSIBLE MOISTURE LEAKAGE
When installing the ECG connector cable, use a
new seal (