LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Installing the System (A01)/Therapy (A04) PCB Assembly
250
8
Replacing the Therapy PCB (A04)
To replace the therapy PCB (A04):
NOTE:
When replacing the therapy PCB, use the
Therapy PCBA Repair Kit (MIN 3305431-008) (p. 502)
.
1.
Place the EMI shield (
) on the therapy PCB and insert using three new screws (
); torque to 6.8 in-lb.
2.
Place system PCB facedown on the work surface with the system/therapy mounting bracket installed.
3.
Verify that the 30 pin header has no bent pins, and insert the 30 pin header (W03) (
) into the J50 circuit side of
the therapy PCB.
4.
Place the spacer (
) into hole in system/therapy mounting bracket with small end of spacer facing up.